If
you do not check and maintain the equipment after the surgery shadowless light is bought, it will cause
frequent failures, makes its life greatly reduced. So how to extend the life of
LED
shadowless lamp?
1.
Periodic inspection, once a year. Check the connection status of each
connection part of the LED
medical light, prevent accidents caused by loose parts. After the operation is
completed, the power must be turned off
2.
To change the fuse, you must turn off the power. Replace the fuse of the same
model. When changing the lamp bulb, you must first cut off the power supply and
wait until the lamp bulb has cooled before replacing it. When the filter glass
is damaged, the factory should be promptly repaired, otherwise it may cause
burns of the body tissue.
3.
Replace the lamp bulb, first cut off the power, then remove the disinfection
handle, and then use a screwdriver to remove the light source system. The
damaged bulb can only be replaced after it has been determined that the lamp
has cooled down.
Note:
Do not handle the bulb directly with your hand to avoid fingerprints on the new
bulb and affect the light source.
4.
Before the operation, the disinfection handle should be removed and disinfected
with 75% alcohol solution. Wipe the lamp head with a soft cloth dampened with
75% alcohol.
Note:
LED
operation light handles, disinfection handles, and light panels must not
be scrubbed with organic solvents.
5.
Balance adjustment. First remove the plastic cover of the balancer which is near
the lamp end and find the round nut inside the balancer tube. Adjusting the up
and down movement force of the lamp head with a special adjustment lever, so
that the movement force of the lamp head up and down is basically the same.
6.
Damping adjustment. Adjustment of the rotational damping of the lamp body
around the main shaft, using a flat-blade screwdriver to adjust, clockwise
rotation.